New Life Radiology presents the most advanced CBCT technology – CONE BEAM COMPUTED TOMOGRAPHY – the biomedical imaging technique for the acquisition of volumetric images of the dental arches. CBCT System represents a valid support for the realization of interventions in implantology, general/maxillofacial surgery, periodontology, endodontics and ATM.
Flat panel CMOS sensor 13 x 13 cm active area with 100nm pixel size
Available real FOV 9×9, 11×11, MULTIFOV (9×9, 11×11, 6×11, 5×5) (not stitching)
Newly developed geometric calibration
Post-processing function for 2D images obtained with 3D sensor
An efficient two-in-one solution to obtain 2D 13 x 30 cm and Volumetric 9 x 9 (or 11 x11 or MULTIFOV 9×9, 11×11, 6×11, 5×5 cm) images in just 10 seconds.
A flat panel CMOS sensor with an active area of 13 x 13 cm, a resolution of 100 microns and an acquisition capacity of 300 frames: with these special features, the acquisition of an ideal image database for the volumetric construction of an image with FOV 9 cm (diameter) x 9 cm (height) is guaranteed (or 11×11, or MULTIFOV 9×9, 11×11, 6×11, 5×5, depending on the chosen configuration)
A unique tool to assist you in implant surgery:
XELIS features a simple interface that helps assess numerous clinical pathologies including fractures, impacted teeth, periodontitis and TMJ.
Xelis
advanced implant DBM
(Axial, Panoramic, Cross Section)
DLB – Dynamic Light Box
Image stitching
Report – Captured Image
Management and Report generation
DICOM Print and CD burning
Net environment, optional multi user up to 10 users
Xelis
basic implant DBM
3D PANORAMIC
Sensor type CMOS
Sensor area 13 x 13 cm
Exposure time 14.3/15.0 sec (Child/Adult, standard PAN)
IMAGE PROGRAMS PAN Adult
panoramic Child panoramic
3-layer focal option PAN
TMJ closed/open mouth
Sinus
Sector panoramic
– Emi panoramic R
– Emi panoramic L
– Low dose pan (optional)
– Ortho panoramic (optional)
– Incisors
– Bitewing R – Bitewing L – Bitewing r + L (optional)
Patient selection: Adult/Child, 3 sizes for all modes
3D IMAGE
Imaging modes Dentition, TMJ R, TMJ L
Field of view 9 x 9 cm (height x diameter) (or 11×11, or MULTIFOV 9×9, 11×11, 6×11, 5×5, depending on the chosen configuration)
Detector pixel size 100µm (200µm in 2×2 binning)
Voxel size 121 µm
Acquisition rate 2 frames per degree
Tube head rotation 230
Dynamic range 14 bit grey level (max 16,384)
Number of frames acquired 460
Scan time/exposure time 12.2 sec/8.5 sec
CEPHALOMETRIC IMAGING
Sensor type DIGITAL FLAT PANEL DR Amorphous Silicon
Image format 24cm x 30cm maximum
Exposure type Single shot DR
Acquisition time Immediate
Setup and exposure time 200ms-500ms
CEPH PROGRAMS L/L
P/A
A/P
Carpus projection
Generator type High frequency DC
Focal point 0.5 mm
Total filtration > 2.5 mm Aleq @ 70 kV)
Stray radiation According to IEC 60601-2-63:
Anode voltage from 61 to 85 kV, step 3 kV
Anode current from 4 to 10mA 9 steps
Weight 110Kg
Dimensions (HxWxD) 2230mm x 1720mm x 1070mm